| Sizes |
|
4, 6, 14, and 21 μm (ISO MTD) |
| Reproducibility |
|
0.5 ISO code (minimum concentration ISO MTD 2.8 mg/L , maximum ISO code = 24) |
| Display |
|
Optional local display presents ISO codes, HIAC codes and alarms |
| Power |
|
9 to 36 VDC, 150 mA (power must be supplied to instrument for operation) |
| Output |
|
RS-232, RS-485, 0 to 5 volts, Modbus, alarms, local and remote displays |
| Reports |
|
ISO 4406 and codes for 4, 6, 14, 21 μm (ISO MTD), particles/mL |
| Fitting Connections |
|
SAE –4 and –8 |
| Sensor Flow Rate |
|
50 to 500 mL/min (0.01 to 0.1 gal/min) through view area; All units except the PT4000-3 offer integrated flow rate controllers |
| Fluid Compatibility |
|
Hydraulic and lubrication oils, mineral, synthetic (Phosphate Ester option) |
| Viscosity |
|
2 cSt (32 SUS) to 424 cSt (viscosities tested at ambient temperature: 25°C +/- 2 degrees) |
| Environment |
|
Operating Temperature -10 to 60°C (-14 to 140°F) |
|
Storage Temperature -40 to 85°C (-40 to 185°F) |
|
97% relative humidity, non-condensing |
| IP |
|
Rating 67 |
| Model PT4000-x |
|
Internal Flow Control Inhibitor |
| Particle Transducer |
-1 |
High pressure 27.6 to 482.5 bar (400 to 7,000 psi) |
| -4 |
High pressure 27.6 to 482.5 bar (400 to 7,000 psi) with local display |
| -7 |
Moderate pressure 3.4 to 82.7 bar (100 to 1200 psi) |
| -8 |
Moderate pressure 3.4 to 82.7 bar (100 to 1200 psi) with local display |
| -9 |
Low pressure 1.4 to 6.9 bar (20 to 200 psi) |
| -0 |
Low pressure 1.4 to 6.9 bar (20 to 200 psi) with local display |
|
Internal Check Flow Control |
| -2 |
3.8 to 37.9 L/min (1 to 10 gal/min), 1.4 to 482.5 bar (20 to 7000 psi) |
| -5 |
3.8 to 37.9 L/min (1 to 10 gal/min), 1.4 to 482.5 bar (20 to 7000 psi) with local display |
|
No Integrated Flow Control |
| -3 |
50 to 500 mL/min (0.01 to 0.1 gal/min), 1.4 to 482.5 bar (20 to 7,000 psi) |
| -6 |
50 to 500 mL/min (0.01 to 0.1 gal/min), 1.4 to 482.5 bar (20 to 7,000 psi) with local display |
| Model IM4000-x Computer/PLC Interface |
-1 |
RS-232 dongle |
| -3 |
Modbus, RS-232/RS-485, 0 to 5 volt |
| -4 |
Remote LED display with programmable contamination alarm levels, dry contact closure (triggers on alarm) |