| Number of Channels |
8 |
| Size Channels |
ISO-MTD (standard) 4, 4.6, 6, 9.8, 14, 21.2, 38, 68 μm |
| ACFTD (optional) ~1, 2, 5, 10, 15, 25, 50, 100 μm |
| Flow Rate |
50 mL/min standard (consult factory for optional offerings down to 15 mL/min) |
| Light Source |
Laser diode |
| Calibration |
ISO MTD (based on ISO 11171) |
| Full ISO 11171 or ISO 4402 optional |
| Counting Efficiency |
Meets JIS B9925:1997 |
| Concentration Limit |
20,000 particles/mL at 5% coincidence loss (per ISO 11171) |
| Sample Volume |
3 runs (averaged) of 5, 10 or 20 mL (programmable) |
| Fluid Temp Range |
0 to 90°C at 25°C ambient (32 to 194°F at 77°F ambient) |
| Measured Fluid Temperature |
0 to 100°C, ±0.5°C (32 to 212°F, ±0.9°F) |
| Viscosity Range |
10 to 424 cSt |
| Measurement |
10 to 424 cSt ±20% at value |
| Wetted Materials |
Aluminum, stainless steel, sapphire, PTFE and Aflas® |
| Cleanliness Classification |
ISO 4406-1991, ISO 4406.2-1999, NAS 1638, MIL-STD-1246C, NAVAIR 01-1A-1, SAE AS 4059 |
| Data Storage |
500 Sample Records |
| Dimensions |
17.8 D x 33.0 W x 35.6 H cm (7 x 12.5 x 14 inches) |
| Weight |
9.5 kg (21 lbs) |
| Input/Output |
Serial Communication RS-232 |
| Bottle Operation |
Purge Volume: 15 to 30 mL programmable |
| Cartridge: CO2, replaceable, rechargeable |
| Operating Capacity: 60 samples per cartridge (120 mL sample bottle) |
| Shop Air: 60 to 110 psi (4.1 to 7.6 bar) clean, dry |
| Online Operation |
Fluid Pressure 40 to 6000 psi (2.75 to 413.7 bar) |
| Purge Volume 15 to 999 mL programmable |
| Power |
DC Input +24 VDC, 2A |
| AC/Battery Adapter Universal 100 to 240 VAC, 50 to 60 Hz, 60 W |
| Rechargeable Battery Nickel-Metal Hydride |
| Operating Time 100 samples or 4 hours continuous |
| Recharge Time 2.5 hours |
| Environment |
Ambient Temperature: 0 to 50°C (32 to 122°F); 20 to 85% relative humidity, non-condensing |
| Storage: -40 to 70°C (-40 to 158°F), up to 98% relative humidity, non-condensing |