| Size Channels (μ) | 2 ch. 0.3, 0.5 |
| 4 ch. 0.3, 0.5, 1.0, 5.0 | |
| 5 ch. 0.3. 0.5, 1.0, 5.0, 10.0 | |
| 6 ch. 0.3, 0.5, 1.0, 3.0, 5.0. 10.0 | |
| Flow Rate | 1.0 CFM (28.3 LPM) ±10% |
| Flow Control | Electronic meter with rear panel adjustment |
| Zero Count | Less than 1 count every 5 minutes |
| Display | Red LED, 7 digits |
| Light Source | Laser diode (10-year MTTF) |
| Outputs | Built-in printer; RS-232/RS-485 serial interface |
| Sample and Hold Times | 1 second to 24 hours |
| Alarm Counts | 0 to 9,999,999 |
| Location Labels | 0 to 999, appear on printout |
| Count Data | Total counts, counts/ft3, counts/m3 |
| Data Storage | 400 samples |
| Coincidence Loss | Less than 5% at 400,000 counts/ft3 |
| FS209E/ISO 14644-1 Calculations | Mean of averages; standard deviation; standard error; UCL |
| Power | 100 VAC, 115 VAC or 230 VAC; 110 watts |
| Dimensions | 28.4 W x 15.2 H x 45.7 D cm (11.2” x 6.0” x 18”) |
| Weight | 10.9 kg (24 lbs) |
| Environment | Operating: 12 to 29°C (55 to 84°F), 20 to 95% relative humidity, non-condensing |
| Storage: -23 to 70°C (-10 to 160°F), up to 98% relative humidity, non-condensing |
| Name | Type | Language | Date |
|---|---|---|---|
| MET ONE-Portable-Air-Particle-Counter-2400 Series-Brochure-English | English | 22 Nov 2011 | |
| MET ONE-Portable-Air-Particle-Counter-2400 Series-Manual-English | English | 22 Nov 2011 |
| Part Number | Description |
|---|---|
| 203478-1 | Isokinetic Probe, w/ tripod and tubing |
| 2080443 | Purge Filter assembly 0.2 µm |
| 460519 | Printer Paper (2 rolls included) (price per roll) |
| 400105 | Power Cord |
| 701087 | Manual, Owners Operating |
| Part Number | Description |
|---|---|
| 460519 | Printer Paper |
| 2080200GP | Filter Scanning Probe w/electronics (Sonalert & LED indicator) 1 CFM |
| 2080200-1 | Filter Scanning Probe w/o electronics 1 CFM |
| 2080732-3 | High Pressure Diffuser, 1.0 CFM (1/4” tube) |
| 826172 | Air Velocity Probe, 0 – 200 FPM |
| 430002-01 | Shipping case (with wheels) |
| 2084045-02 | Particle Vision PortAll, version 2.X, Kit, download utility software, adapter and manual (serial cable must be purchased separately) |
| 400130 | Cable, Serial Communication, DB-9 male to DB-9 male |
| 700011-21 | PortAll License Standard edition |
| 700011-22 | PortAll License 21 CFR Part 11 compliant |
| 701169-01 | PortAll Validation Documentation (IQ/OQ) CD |
| Q. | Can the MET ONE Model 2400 be used with a manifold? |
| A. | An ultra low cost site monitoring solution can be deployed when the MET ONE Model 2400 is coupled with the MET ONE Model A3432 manifold. Sample sequentially from up to 32 locations with a single instrument. |
| Q. | Can the MET ONE 2400 be used for HEPA filter scanning? |
| A. | Yes, the MET ONE 2400 is ideally suited to this application. HEPA leaks are easy to detect when used in conjunction with the 1.0 CFM filter scanning probe (part number 2080200-1) with audio and visual feedback on discrete particle events. |
| Q. | Can the MET ONE 2400 be calibrated to 21501-4? |
| A. | Hach does not offer ISO 21501-4 calibrations for the MET ONE 2400 product. Customers requiring the performance benefits of ISO21501-4 should consider the MET ONE 3400 Series. |