| Minimum threshold Setting | R4803 / R4903 - 0.3um R4805 / R4815/ R4905 / R4915 - 0.5um |
| Flow Rate | R4803 / R4805 / R4903 / R4905 - 0.1 CFM (2.83 LPM) R4815 / R4915 - 1.0 CFM (28.3 LPM) |
| Flow Control | Critical orifice; requires 18” Hg vacuum, minimum 450 mbar |
| Size Channels (μm) | R4803 / R4903 - Ch 1 / Ch2 - 0.3 / 0.5 R4805 / R4815 / R4905 / R4915 - Ch 1 / Ch2 - 0.5 / 5.0 |
| Light Source | Laser diode (10-year MttF) |
| Coincidence Loss | 5% at 2,000,000 particles/ft3 |
| Inlet Pressure | Ambient to 0.1” Hg vacuum |
| Indicators | Power LED illuminates when counter is powered; Calibration/Alarms LED illuminates when sensor calibration needs checking and flashes when the count exceeds the alarm level. |
| Power | Open collector Fet maximum sink current (on) 150 mA; Maximum VC voltage (off) 30 VDC |
| Connector | DB-15 (female) |
| Weight | 0.3 kg (10.7 oz) |
| Tubing Connections | Inlet, 0.1 CFM, 1/8” I.D. Inlet, 1.0 CFM, 1/4” I.D. Outlet, All (vacuum), 1/4” I.D. |
| Environment | Operating - 12to 41C (55 to 105F) 20 to 95% relative humidity, non-condensing Storage - -40 to 70C (-40 to 160F) up to 98% relative humidity, non-condensing |
| When ordering, specify | 0.3 µm (R4803/R4903) or 0.5 µm (R4805/R4815/R4905/R4915) AND 0.1 cfm (R4803/R4805/R4903/R4905) or 1.0 cfm (R4815/R4915) |
| Name | Type | Language | Date |
|---|---|---|---|
| MET ONE-Remote-Air-Particle-Counter-4800 Series-Brochure-English | English | 22 Nov 2011 | |
| MET ONE-Remote-Air-Particle-Counter-4900 Series-Brochure-English | English | 22 Nov 2011 | |
| MET ONE-Remote-Air-Particle-Counter-4800 Series-Manual-English | English | 22 Nov 2011 | |
| MET ONE-Remote-Air-Particle-Counter-4900 Series-Manual-English | English | 22 Nov 2011 | |
| Proven-Long-Life-Lasers-in-Optical-Particle-Counters | English | 22 Nov 2011 | |
| MET ONE-Remote-Air-Particle-Counter-4800 Series-Brochure-UK | English A4 | 22 Nov 2011 | |
| MET ONE-Remote-Air-Particle-Counter-4900 Series-Brochure-UK | English A4 | 22 Nov 2011 |
| Part Number | Description |
|---|---|
| 500052 | Isokinetic probe; DB-15 Connector |
| 701113 | Operators Manual R4800 |
| 701130 | Operators Manual R4900 |
| Part Number | Description |
|---|---|
| 2080442 | Purge Filter Assembly, 0.2µ , 0.1 cfm (R4803 / R4805 / R4903 / R4905) |
| 2080443 | Purge Filter Assembly, 0.2µ , 1.0 cfm (R4815/R4915) |
| 2082270-1 | Power Supply, 110 VAC input, 12 vdc output supports up to 32 model R4800 counters |
| 2082270-2 | Power Supply, 230 VAC input, 12 vdc output supports up to 32 model R4800 counters |
| 2081524-11 | Wall Plate, R4800/R4900 Series |